3. OPERATION 
 
Page 47 
© 2007 DH Instruments, a Fluke Company 
3.4 
[SETUP] 
 PURPOSE 
[SETUP] accesses a menu of functions and features commonly used in setting up and using RPM4. 
 OPERATION 
To access the SETUP menu, press [SETUP] from the MAIN RUN 
screen.  The display is: 
1range 2res 3stab  
4UL 5ATest 
 
SETUP menu choices include: 
<1range>  
Save and delete ranges created by AutoRange (see Section 3.4.1). 
<2res> 
Adjust the resolution of pressure displays (see Section 3.4.2). 
<3stab> 
Adjust the pressure stability limit that is the criterion for the Ready/Not Ready indication 
(see Section 3.4.3). 
<4UL> 
Adjust upper pressure limit alarm (see Section 3.4.4). 
<5ATest> 
Set up File AutoTest files and view data files resulting from running AutoTests (See 
Section 3.4.5). 
3.4.1 
<1RANGE> 
 PURPOSE 
Save AutoRange ranges and associated settings for recall.  Delete previously saved AutoRange ranges. 
 PRINCIPLE 
The RPM4 AutoRange function creates a range and automatically sets various operating 
parameters proportionally to the range (see Section 3.3.4).  Operating settings can then be 
changed by the user.  Many operating settings are specific to the current range (see Table 2). 
[SPECIAL]<1range> allows the AutoRange range and associated settings to be saved for 
recall using the [RANGE] key (see Section 3.3.1).  This can avoid having to recreate 
frequently used ranges and settings.  
[SPECIAL]<1range> is also used to delete ranges that have been saved but are no longer needed. 
3.4.1.1 
SAVING AN A
UTO
R
ANGE
 RANGE 
 OPERATION 
To save a range created by AutoRange, the range must be the active range.   
Use AutoRange or AutoTest to create the range (sees Sections 3.3.4, 3.3.10) 
and then make any desired feature and setting adjustments. 
Once the AutoRange range is set up as desired, press [SETUP], <1range>
<1save>.   The display is: 
1. 
Indication that this is a range to be saved. 
2. 
Designator and position of the Q-RPT used by 
the range to be saved. 
3. 
Pressure unit of measure of the range to be 
saved.  
4. 
Full scale pressure of the range to be saved in 
gauge mode and absolute mode if available. 
    
                    
 
Save range A70M   Hi 
psi   4000g/4000a