Omega DPF78 User Manual

Page of 30
12
ENTER
STEP
3
REMARKS
PRESS
DISPLAY
This section of the menu is used to set up the
rate information.
Press the PRGM key to choose RPS (rate per
second) or SCALE (RPM, RPH). Press ENTER
to enter displayed choice.
Press PRGM to choose ÷60 (RPM) or ÷3600
(RPH). Press ENTER to enter displayed choice.
This sets the normalizing (averaging) factor.
Press the arrow keys under the desired digits to
change. Press ENTER to enter displayed value.
This sets the number of significant figures to be
displayed. Press the arrow key under the digit to
change. Press ENTER to enter displayed value.
This sets the delay time (0.0 to 9.9) that the unit
will "look" for valid input data before the display
falls to 0. Press the arrow key under the digits to
change. Press ENTER to enter displayed value.
factor
count
rate
rps
or
scale
z 60
or
z 3600
nor ##
figur #
dly #.#
ENTER
ENTER
PRGM
PRGM
PRGM
ENTER
ENTER
SETTING
THE RATE
DPF75
&
DPF78
ONLY