National Instruments NI 4070/4072 6 1/2-Digit FlexDMM User Manual
NI 4070/4072 Calibration Procedure
22
ni.com
31. Reset the calibrator for safety reasons.
You have completed verifying the AC voltage of the NI 4070/4072. Select
one of the following options:
one of the following options:
•
If you want to continue verifying other modes, go to the Verifying
4-Wire Resistance
4-Wire Resistance
•
If you do not want to verify other modes and you are performing a
post-adjustment verification, go to the
post-adjustment verification, go to the
section.
•
If you do not want to verify any additional modes and you are
performing a pre-adjustment verification, call
performing a pre-adjustment verification, call
niDMM_close
to close
the session.
Verifying 4-Wire Resistance
To verify the 4-wire resistance of the NI 4070/4072, complete the
following steps:
following steps:
1.
Reset the calibrator.
2.
Fasten the connectors on one end of each Fluke 5440 cable to the
appropriate banana plug connectors on the NI 4070/4072. Fasten the
connectors on the other end of each Fluke 5440 cable to the appropriate
calibrator binding posts. Figure 2 shows the Fluke 5440 cables. Table 7
lists the cable connections.
appropriate banana plug connectors on the NI 4070/4072. Fasten the
connectors on the other end of each Fluke 5440 cable to the appropriate
calibrator binding posts. Figure 2 shows the Fluke 5440 cables. Table 7
lists the cable connections.
Figure 2. Cable Connections for 4-Wire Resistance
1
NI 4070/4072
2
Fluke 5700A/5720A Calibrator
3
Fluke 5440 Cables
HI
LO
300V
MAX
HI
LO
AUX
I/O
1A, 250V
MAX
300V
MAX
5V
MAX
AMPS
INPUT
SENSE
W 4W
6½-Digit FlexDMM
CAT II
V
W
HI
LO
HI
LO
HI
SENSE
V
Ω
OUTPUT
V
ΩA
AUX
CURRENT
GUARD
GROUND
1
2
3