NI 4070/4072 Calibration Procedure
22
ni.com
31. Reset the calibrator for safety reasons.
You have completed verifying the AC voltage of the NI 4070/4072. Select 
one of the following options:
If you want to continue verifying other modes, go to thVerifying 
4-Wire Resistance
 section.
If you do not want to verify other modes and you are performing a 
post-adjustment verification, go to the 
 section. 
If you do not want to verify any additional modes and you are 
performing a pre-adjustment verification, call 
niDMM_close
 to close 
the session.
Verifying 4-Wire Resistance
To verify the 4-wire resistance of the NI 4070/4072, complete the 
following steps:
1.
Reset the calibrator.
2.
Fasten the connectors on one end of each Fluke 5440 cable to the 
appropriate banana plug connectors on the NI 4070/4072. Fasten the 
connectors on the other end of each Fluke 5440 cable to the appropriate 
calibrator binding posts. Figure 2 shows the Fluke 5440 cables. Table 7 
lists the cable connections.
Figure 2.  Cable Connections for 4-Wire Resistance
1
NI 4070/4072
2
Fluke 5700A/5720A Calibrator
3
Fluke 5440 Cables
HI
LO
300V
MAX
HI
LO
AUX
I/O
1A, 250V
MAX
300V
MAX
5V
MAX
AMPS
INPUT 
SENSE
W 4W
6½-Digit FlexDMM
CAT II
 
V
W
HI
LO
HI
LO
HI
SENSE
V
Ω
OUTPUT
V
ΩA
AUX
CURRENT
GUARD
GROUND
1
2
3