NI 4070/4072 Calibration Procedure
3.
Wait two minutes for the thermal EMF to stabilize if the Fluke 5440 
cables were not previously connected in this configuration. 
4.
Call 
niDMM_reset
.
5.
Refer to Table 8 for the appropriate calibrator output and function 
parameter values as you complete the following steps:
a.
On the calibrator, output the value listed in the Calibrator Output 
column in Table 8 for the current iteration. Make sure that the 
external sense is turned on but 2-wire compensation is turned off.
Note
After setting the calibrator output to 0
Ω in the seventh iteration, you do not need to 
continually set the calibrator to 0
Ω for iterations 8 through 12.
b.
Call 
niDMM_ConfigureMeasurement
 with the parameters set 
as shown in Table 8 for the current iteration.
c.
Call 
niDMM_ConfigureOffsetCompOhms
 with 
OffsetCompOhms set to either 
NIDMM_VAL_OFFSET_COMP_OHMS_ON
 or 
NIDMM_VAL_OFFSET_COMP_OHMS_OFF
 according to Table 8 for 
the current iteration.
d.
Call 
niDMM_Read
. Verify that this measurement falls between the 
tolerances listed in Table 17. Tolerances are provided instead of 
absolute limits because your calibrator will have different discrete 
resistance values.
Table 7.  Fluke 5440 Cable Connections
Fluke 5440 
Cable Identification
Banana Plug Connector
(NI 4070/4072)
Banana Plug Color
(Fluke 5440 Cable)
Binding Post
(Fluke 5700A/5720A Calibrator)
First cable
HI
Red
OUTPUT HI
LO
Black
OUTPUT LO
(No connection)
Blue
V GUARD
Second cable
HI SENSE
Red
SENSE HI
LO SENSE
Black
SENSE LO
(No connection)
Blue
V GUARD