National Instruments NI 4070/4072 6 1/2-Digit FlexDMM User Manual

Page of 68
© National Instruments Corporation
25
NI 4070/4072 Calibration Procedure
4.
Call 
niDMM_ConfigureMeasurement
 with the following 
parameters:
Function = 
NIDMM_VAL_2_WIRE_RES
Range = 
10e3
Resolution = 
0.01
5.
Call 
niDMM_ConfigureOffsetCompOhms
 with OffsetCompOhms 
set to 
NIDMM_VAL_OFFSET_COMP_OHMS_ON
.
6.
Call 
niDMM_Read
. Verify that this measurement falls between the 
limits listed in Table 18.
7.
Call 
niDMM_ConfigureMeasurement
 with the following 
parameters:
Function = 
NIDMM_VAL_2_WIRE_RES
Range = 
1e3
Resolution = 
1e–3
8.
Call 
niDMM_ConfigureOffsetCompOhms
 with OffsetCompOhms 
set to 
NIDMM_VAL_OFFSET_COMP_OHMS_ON
.
9.
Call 
niDMM_Read
. Verify that this measurement falls between the 
limits listed in Table 18.
10. Call 
niDMM_ConfigureMeasurement
 with the following 
parameters:
Function = 
NIDMM_VAL_2_WIRE_RES
Range = 
100
Resolution = 
100e–6
11. Call 
niDMM_ConfigureOffsetCompOhms
 with OffsetCompOhms 
set to 
NIDMM_VAL_OFFSET_COMP_OHMS_ON
.
12. Call 
niDMM_Read
. Verify that this measurement falls between the 
limits listed in Table 18.
13. Remove the shorting bar from the NI 4070/4072.
14. Reset the calibrator.
15. Fasten the connectors on one end of the Fluke 5440 cable to the 
NI 4070/4072, and fasten the connectors on the other end of the cable 
to the appropriate calibrator binding posts. Figure 1 shows the correct 
connections. Table lists the cable connections.
16. Wait two minutes for the thermal EMF to stabilize if the Fluke 5440 
cable was not previously used in this configuration. 
17. Output 0
Ω on the calibrator with 2-wire compensation turned on 
but with external sense turned off.
18. Call 
niDMM_reset
 to reset the NI 4070/4072 to a known state.