Trendnet 18-1106-33 User Manual

Page of 27
4. Operation
the external system and the level in the bath will fall. Add
more liquid until the level is once again 2 cm below the top
plate. 
If the safety circuit is triggered, please top up with more liqu-
id and restart the unit using 
. Repeat this action if
necessary.
Note: Check the liquid level after the unit has been in use
for a while.
When MultiTemp III is switched on, the compressor starts
and the heater will be activated if the set temperature is grea-
ter then the actual temperature. A constantly lit dot in the
left digit position (9 in Fig. 2) indicates that the heater is
working to reach the set temperature. A flashing dot indica-
tes that the bath has reached the set temperature.
Note: Setting a temperature limit is important for the safe
operation of the unit. It does not set the working tem-
perature.
Set the temperature limit using the dial on the
rear of the unit. The scale shows approximate
temperatures. Set the dial to a value 5-10 °C
above the desired working temperature. 
If a more accurate temperature limit is required, the procedu-
re described in Section 7.4 should be followed.
The range of temperatures that can be set can be limited if
required by the application or for other reasons. 
Caution: This is a setting limitation, not a primary safety
measure. The temperature limit protection must be set sepa-
rately (see Section 4.2).
1. Press          to select the H mode 
10
4.2 Setting the
temperature
limit
4.3 Tempera-
ture setting 
limit