A&D AD-4212B-101/102/201/301 User Manual

Page of 102
 
25
 
 
5  Confirm that there is nothing on the pan and press the 
PRINT key. The balance measures the zero point and 
displays the measured value. Do not allow vibration or 
drafts to affect the balance. 
  The balance displays the calibration weight value. 
6  Place a calibration weight, of the weight value displayed, 
on the pan and press the PRINT key. The balance 
measures the calibration weight and displays the 
measured value. Do not allow vibration or drafts to affect 
the balance. 
7  The balance displays    end . Remove the weight from the 
pan. 
8 The balance displays  glp    and outputs “Calibration Test 
Report” using the RS-232C interface or stores the 
calibration test data in memory. For details on the 
calibration test report format, refer to “11-2 GLP Report”. 
9 The balance will automatically return to the weighing 
mode. 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
Calibration
weight
GLP output
 From previous page