9.  FUNCTION SWITCH AND INITIALIZATION 
9-1  Permit or Inhibit 
The balance stores parameters that must not be changed unintentionally (e.g. Calibration data for 
accurate weighing, data for adapting to the operating environment, control data for the RS-232C 
interface). There are two switches for protecting the parameters. The switches can select either 
“permit” or “inhibit”. The “inhibit” protects parameters against unintentional operations. There is an 
additional switch for the extended functions of the AD-4212A. 
 
Switches 
 
 
 
 
 
 
 
 
Note: Only available for the AD-4212A series balance. 
For details, refer to “20. EXTENDED FUNCTIONS 
 
 
 
 
 
 
Operation 
1  Press the ON:OFF key to turn off the display. 
2  While pressing and holding the PRINT key and the SAMPLE key, press the ON:OFF key. The 
balance displays    p5  . 
3  Press the PRINT key. Then the balance displays the function switches. 
4  Set the switches using the following keys. 
SAMPLE key 
To select the switch to change the parameter. 
RE-ZERO key 
To change the parameter of the switch selected. 
 
0: To inhibit changes.      1: To permit changes 
PRINT key 
To store the new parameter and return to the weighing mode.   
CAL key 
To cancel the operation and return to the weighing mode.