A&D AD-4212B-101/102/201/301 User Manual

Page of 102
 
80
 
 
Display             Error code 
Description 
 
Unit mass error 
The sample unit mass for the counting mode is too 
light. Storing and using it for counting will cause a 
counting error. 
Add samples to reach the specified number and press 
the PRINT key. 
Pressing the PRINT key without adding samples will 
shift the balance to the counting mode. But, to acquire 
accurate weighing, be sure to add samples. 
 Clock 
battery 
error 
The clock backup battery has been depleted. 
Press any key and set the time and date. The clock and 
calendar function works normally as long as the AC 
adapter is connected to the balance. If this error appears 
frequently, contact the local A&D dealer. 
 
                          (Blinking) 
Memory full 
 The amount of weighing data in memory has reached 
the maximum capacity. 
 Delete data in memory to store new data. For details, 
refer to “14. DATA MEMORY”. 
 
                          
(Illuminated) 
Memory full 
The amount of calibration or calibration test data in 
memory has reached the maximum capacity (50 sets). 
 The data in memory will be deleted automatically to 
store new data. For details, refer to “14. DATA 
MEMORY”. 
 
Memory type error 
Type of memory set in the function table and type of 
data stored are different. 
For details, refer to “14. DATA MEMORY”. 
EC, E00 
Communications error 
A protocol error occurred in communications.   
Confirm the format, baud rate and parity. 
  EC, E01 
Undefined command error 
An undefined command was received.   
Confirm the command. 
EC, E02 
Not ready 
A received command can not be processed.   
e.g.  The balance received a Q command, but not in the 
weighing mode. 
e.g. The balance received a Q command while  
processing a RE-ZERO command. 
Adjust the delay time to transmit a command.