A&D AD-4212B-101/102/201/301 User Manual

Page of 102
 
82
23. SPECIFICATIONS 
 
 
AD-4212A-100 AD-4212A-200 AD-4212A-600
AD-4212A-1000 
Weighing capacity 
110 g 
210 g 
610 g 
1100 g 
Maximum display 
110.0084 g 
210.084 g 
610.084 g 
1100.084 g 
Minimum weighing value (1 digit) 
0.1 mg 
1 mg 
Repeatability (Standard deviation) 
0.15 mg 
1 mg 
Linearity 
±
0.3 mg 
±
2 mg 
±
3 mg 
0-30 g 
1.1
0-30 g
0.8
0-30 g 
0.9
0-30 g 
0.9 
Stabilization time in seconds 
(typical at FAST under good 
environment) 
30-110 g  1.3 30-210 g 1.0 30-610 g 1.1
30-1100 g 
1.1 
Display refresh rate   
10 times/second (5 times/second can be selected) 
RS-232C 
Bi-directional, 600-19200 bps 
Comparator contact 
output 
3-level output 
I/O unit 
External contact input 
RE-ZERO, PRINT 
Buzzer 
Sounds the connected buzzer for 3-level comparison results 
(It becomes 5-level when OP-04 is installed) 
Sensitivity drift   
±2 ppm/°C (10°C-30°C/50
°F-86°F ) 
Operating environment   
5°C to 40°C (41°F to 104°F), 85% RH or less (No condensation) 
Calibration weight provided 
(Conforming to OIML Class E2) 
50  g 
100  g 
200 g 
Applicable weight values 
20 g, 50 g, 100 g
20 g, 50 g,   
100 g, 200 g 
20 g, 50 g, 100 g,
200 g, 300 g, 400 g,
  500 g, 600 g 
20 g, 50 g, 100 g, 200 g, 
300 g, 400 g, 500 g, 600 g,
700 g, 800 g, 900 g, 1000 g
Dimensions 80 
(W) 
×
 230 (D) × 90 (H) mm 
Weighing pan 
50  × 50 mm 
70 × 70 mm 
Weighing 
unit 
Mass Approx. 
kg 
Dimensions 237 
(W) 
×
 150 (D) × 155 (H) mm 
Display 
Mass Approx. 
kg 
Connection cable   
Approx. 2 m (between weighing unit and display) 
AC adapter   
Confirm that the adapter type is correct for the local voltage and power receptacle type 
Power consumption 
Approx. 11VA (supplied to the AC adapter ) 
Minimum unit mass 
0.1 mg 
1 mg 
Counting 
mode 
Number of samples 
10, 25, 50 or 100 pieces 
Minimum 100% 
reference mass 
10.0 mg 
100 mg 
Percent 
mode 
Minimum 100% 
display 
0.01%、0.1%、1% 
(Depends on the reference mass stored.) 
OP-01 
BCD output, external contact input 
RS-232C and comparator contact output disabled) 
OP-04 RS-232C, 
5-level 
comparator 
contact output, external contact input 
(Not field installable. Must be ordered with the balance for installation at the factory.)
OP-07* Extension 
cable 
OP-08 Ethernet 
interface 
 
RS-232C and comparator contact output disabled) 
OP-19 
Stainless steel breeze break  (Provided as standard for AD-4212A-100) 
Options 
OP-20 Metal 
leveling 
foot 
 
Only one extension cable can be used. The maximum distance between the weighing unit 