Fluke 2620A User Manual

Page of 197
  Appendices
Specifications
A
A-13
Isolation
none
dc-coupled
Threshold
1.4V
Hysteresis
500 mV
Trigger Input
Input Voltages
contact closure and TTL compatible
“high” =2.0V min, 7.0V max
“low” = -0.6V min, 0.8V max
Isolation
None
dc-coupled
Minimum Pulse Width
5 µs
Maximum Frequency
5 Hz
Specified Conditions
The instrument must be in the quiescent state, with no interval scans in process, no
commands in the queue, no RS-232 or IEEE interface activity, and no front panel
activity if the latency and repeatability performance is to be achieved. For addition
information, refer to Chapter 5.
Maximum Latency
Latency is measured from the edge of the trigger input to the start of the first channel
measurement for the Specified Conditions (above).
480 ms for fast rate, scanning DCV, ACV, ohms, and frequency only
550 ms for fast rate, scanning any thermocouple or 100 mV dc channels
440 ms for slow rate, scanning DCV, ACV, ohms, and frequency only
890 ms for slow rate, scanning any thermocouple or 100 mV dc channels
Repeatability
3 ms for the Specified Conditions (above)