Specifications
A
A-15
Altitude
Operating: 2,000 m maximum
Non-operating: 12,200 m maximum
Vibration
0.7 g at 15 Hz
1.3 g at 25 Hz
3 g at 55 Hz
Shock
30 g half sine per Mil-T-28800
Bench handling per Mil-T-28800
General
Channel Capacity
21 Analog Inputs
4 Alarm Outputs
8 Digital I/O (inputs/outputs)
Measurement Speed
Slow rate: 4 readings/second nominal
Fast rate: 17 readings/second nominal
1.5 readings/second nominal for ACV and high-
Ω
 inputs
For additional information, refer to Typical Scanning Rated and Maximum Autoranging
Time.
Memory Life
10 years minimum over Operating Temperature range
Stores: real-time clock, set-up configuration, and measurement data
Common Mode Voltage
300V dc or ac rms maximum from any analog input (channel) to earth provided that
channel to channel maximum voltage ratings are observed.
Voltage Ratings
Channels 0,1, and 11 are rated at 300V dc or ac rms maximum from a channel terminal
to earth and from a channel terminal to any other channel terminal.
Channels 2 to 10 and 12 to 20 are rated at 150V dc or ac rms maximum from a channel
terminal to any other channel terminal within channels 2 to 10 and 12 to 20.
IEC Overvoltage Category II.
Size
9.3 cm high, 21.6 cm wide, 31.2 cm deep
Weight
Net, 2.95 kg
Shipping, 4.0 kg