Fluke 1744 User Manual

Page of 73
1744/1743 
Users Manual 
14 
3.  Run the PQ Log software as described in the PQ Log Users Manual. 
4.  Set up the Logging job and transfer the settings to the Logger. 
Test Leads – Markings 
The 1744/1743 Logger includes built-in, labeled test leads for voltage 
terminals L1 or A, L2 or B, L3 or C, and N, as well as two for the internal 
power supply. The Flexi Set or current clamp sets are connected by a seven-pin 
plug to the Logger. Color coding clips are provided for your convenience.  
Table 5. Test Lead Markings 
Test Leads 
Markings 
Phase L1 or A 
L1 / A 
Phase L2 or B 
L2 / B 
Phase L3 or C 
L3 / C 
Neutral wire N 
Supply “Supply” 
Supply “Supply” 
Connecting Current Probes 
Connect current clamps and Flexi Set probes so that current will flow in the 
direction marked by arrows on the probes. Current must flow from the energy 
generator to the energy consumer (the load) in order to maintain a positive 
active power. (The polarization of the test lead for neutral conductor current is 
not significant, because the phase angle of the neutral conductor current is not 
evaluated.)  
Note 
Make sure the clip-on probes are connected to the appropriate phase: 
V
L1
 with I
L1
 for a P-N measurement or V
L12
 with I
L1
 for a P-P 
measurement. 