Specifications
84
AFG3000 Series Quick Start User Manual
Electrical (AFG3011)
Operating Mode
Run Mode
Continuous, Modulation, Sweep, and Burst
Burst Count
1 to 1,000,000 cycles or Infinite
Internal Trigger Rate
1.000 
μs to 500.0 s
Waveforms
Standard
Sine, Square, Pulse, Ramp, More (Sin(x)/x, Noise, DC, Gaussian, Lorentz, 
Exponential Rise, Exponential Decay, and Haversine)
Arbitrary Waveform
Waveform Length
2 to 131,072
Sampling Rate
250 MS/s
Resolution
14 bits
Non-volatile Waveform Memory
4
Frequency
Sine
1
μHz to 10 MHz
Square
μHz to 5 MHz
Pulse
1 mHz to 5 MHz
Ramp, Sin(x)/x, Gaussian, Lorentz, 
Exponential Rise, Exponential Decay,
Haversine
μHz to 100 kHz
Arbitrary
2
1 mHz to 5 MHz
Resolution
μHz or 12 digits
✔Accuracy (Stability)
±1 ppm, 0 
°C to 50 °C (except Arb)
±1 ppm ±1 
μHz, 0 °C to 50 °C (Arb)
Accuracy (Aging)
±1 ppm/year
Phase (except DC, Noise, Pulse)
Range
3
–180.00
° to +180.00°
Lead Delay (Pulse)
Range (Continuous Mode)
0 ps to Period
Range (Triggered/Gated Burst Mode)
0 ps to Period – [Pulse Width + 0.8 * (Leading Edge Time + Trailing Edge 
Time)]
Resolution
10 ps or 8 digits