Tektronix AFG3000 Series Manual De Usuario

Descargar
Página de 108
Specifications
84
AFG3000 Series Quick Start User Manual
Electrical (AFG3011)
Operating Mode
Run Mode
Continuous, Modulation, Sweep, and Burst
Burst Count
1 to 1,000,000 cycles or Infinite
Internal Trigger Rate
1.000 
μs to 500.0 s
Waveforms
Standard
Sine, Square, Pulse, Ramp, More (Sin(x)/x, Noise, DC, Gaussian, Lorentz, 
Exponential Rise, Exponential Decay, and Haversine)
Arbitrary Waveform
Waveform Length
2 to 131,072
Sampling Rate
250 MS/s
Resolution
14 bits
Non-volatile Waveform Memory
4
Frequency
Sine
1
μHz to 10 MHz
Square
μHz to 5 MHz
Pulse
1 mHz to 5 MHz
Ramp, Sin(x)/x, Gaussian, Lorentz, 
Exponential Rise, Exponential Decay,
Haversine
μHz to 100 kHz
Arbitrary
2
1 mHz to 5 MHz
Resolution
μHz or 12 digits
✔Accuracy (Stability)
±1 ppm, 0 
°C to 50 °C (except Arb)
±1 ppm ±1 
μHz, 0 °C to 50 °C (Arb)
Accuracy (Aging)
±1 ppm/year
Phase (except DC, Noise, Pulse)
Range
3
–180.00
° to +180.00°
Lead Delay (Pulse)
Range (Continuous Mode)
0 ps to Period
Range (Triggered/Gated Burst Mode)
0 ps to Period – [Pulse Width + 0.8 * (Leading Edge Time + Trailing Edge 
Time)]
Resolution
10 ps or 8 digits