Princeton 4300 User Manual

Page of 112
Chapter 4 
System Setup 
37 
 
 
 
37 
 
Figure 12.  RSConfig dialog 
4.  Open the WinX application and, from 
Setup|Hardware…,
 run the Hardware 
Setup wizard. 
5.  When the PVCAM dialog (Figure 13) is displayed, click in the 
Yes
 radio button, 
click on 
Next
 and continue through the wizard. After the wizard is finished, the 
Controller/Camera
 tab card will be displayed with the 
Use PVCAM
 checkbox 
selected. You should now be able to set up experiments and acquire data. 
 
Figure 13.  Hardware Setup wizard: PVCAM dialog 
6.  Run the software in focus mode to verify communication between the ST-133 
and the host computer. 
Making the Circulator-Camera Connections 
1.  Set up the coolant circulator according to the directions in the user manual for that 
equipment. Do not apply power to the circulator until directed to do so.  
2.  Make the hose connections between the circulator and the camera. For best cooling 
performance, the tubing should be no longer than necessary.   
  PI-SCX:1300
   Use 1/4", thin-wall plastic tubing. Be sure the tubing is properly 
secured at both ends. Note that the ports on this camera use a ferrule-less quick-
disconnect method of securing the tubing and that both the camera's valve body 
and the fitting insert include automatic shutoff to prevent coolant leaks when 
disconnected. For best cooling, connect the inflow and outflow tubing to the 
ports as indicated in 
Figure 14.