Princeton 4300 User Manual

Page of 112
Chapter 7 
Troubleshooting 
75 
 
 
75 
Detector Temperature, Acquire, and Focus are Grayed Out 
(Versions 2.5.19.0 and earlier) 
These functions and others will be deactivated if you have installed a camera being run 
under USB 2.0 and have opened WinView/32 without having first turned on the ST-133. 
They will also be deactivated if you have installed a camera being run under USB 2.0 and 
a Princeton Instruments high speed PCI card was also detected when RSConfig.exe was 
run. 
1.  Check to see if the ST-133 is connected to the host computer and is turned on. If 
it is not connected or is connected but not turned on, go to Step 2. If it is 
connected and on, go to Step 3.  
2.  Close WinView, verify that the ST-133 is connected to the host computer, turn on 
the ST-133, and reopen WinView. The formerly grayed out functions should now 
be available. 
3.  If the ST-133 is connected and on, the USB 2.0 camera may not be listed as 
Camera 1 in the PVCAM.INI file.  
4.  Run RSConfig.exe (accessible from the Windows|Start|Programs|PI Acton 
menu). If the USB 2.0 camera is listed as Camera 2 (Princeton Style (USB2) in 
Figure 41), you will have to edit the PVCAM.INI file. 
 
Figure 41.  RSConfig dialog: Two Camera Styles 
5.  Using Notepad or a similar text editor, open PVCAM.INI, which is located in 
the Windows directory (C:\Windows, for example).  
If the contents of the file look like:  Change the headings so the contents now look like: 
[Camera_1] 
Type=1 
Name=Camera1 
Driver=rspipci.sys 
Port=0 
 
[Camera_2] 
Type=1 
Name=Camera2 
Driver=apausb.sys 
Port=0 
[Camera_2] 
Type=1 
Name=Camera1 
Driver=rspipci.sys 
Port=0 
 
[Camera_1] 
Type=1 
Name=Camera2 
Driver=apausb.sys 
Port=0