Agilent Technologies E4445A User Manual

Page of 664
Chapter 5
371
One-Button Measurement Functions
Meas Setup (Burst Power)
One-
But
ton
M
e
asur
e
m
e
nt
Funct
ions
5.8
Meas Setup (Burst Power)
When
Burst Power
has been selected in the
Measure
menu of the Spectrum Analysis Mode,
this key displays the appropriate measurement setup menu.
The burst power measurement is an accurate method of determining the average power for
the specified burst.
Key Path:
Front-panel key
History:
Added with firmware revision A.02.00
Remote Command:
There is no equivalent remote command.
5.8.1
Avg Number
Press
Avg Number (On)
to specify the number of measurement averages used when
calculating the measurement result. The average is displayed at the end of each sweep.
Key Path:
Meas Setup
State Saved:
Saved in instrument state.
Factory Preset: 10 averages / Off
Range:
1 to 1000
History:
Added with firmware revision A.02.00
Remote Command:
[:SENSe]:BPOWer:AVERage:COUNt <integer>
[:SENSe]:BPOWer:AVERage:COUNt?
[:SENSe]:BPOWer:AVERage[:STATe] OFF|ON|0|1
[:SENSe]:BPOWer:AVERage[:STATe]?
Example:
BPOW:AVER:COUN 100
BPOW:AVER ON