One-Button Measurement Functions
Meas Setup (Burst Power)
One-
But
ton
M
e
asur
e
m
e
nt
Funct
ions
[:SENSe]:BPOWer:AVERage:TYPe?
Example:
BPOW:AVG:TYP LPOW
er to select Log type.
BPOW:AVG:TYP?
5.8.4
Threshold Lvl
Enables you to set the level above which the mean carrier power calculation is based. The
threshold level can be described in dB (
Rel
) or dBm (
Abs
).
Key Path:
Meas Setup
Factory Preset:
−30 dB
Terminators:
dB or dBm
Default Terminator dB
Knob Increment: 0.1 dB/dBm
Step Key
Increment:
6 dB/dBm
Range:
–60 dBm to 60 dBm (in absolute mode)
–60 dB to 0 dB (in relative mode)
History:
Added with firmware revision A.02.00
Remote Command:
[:SENSe]:BPOWer:THReshold <number>
[:SENSe]:BPOWer:THReshold?
[:SENSe]:BPOWer:THReshold:TYPE ABSolute|RELative
[:SENSe]:BPOWer:THReshold:TYPE?
Example:
SENS:BPOW:THR:TYPE ABS
5.8.5
Meas Method
Allows you to select the measurement method.
Above Threshold Lvl
— Selects the user defined threshold level or default level (-3.00 dB)
as the criteria in making the measurement.
Measured Burst Width
— This measurement method is not available for the following radio
standards: IS-95, J-STD-008, cdma2000-SR1, cdma2000-SR3, W-DCMA 3GPP.