Chapter 5
375
One-Button Measurement Functions
Meas Setup (Burst Power)
One-
But
ton
M
e
asur
e
m
e
nt
Funct
ions
Range:
0.1
µs to 2 ks
History:
Added with firmware revision A.02.00
Remote Command:
[:SENSe]:BPOWer:BURSt:WIDTh <time>
[:SENSe]:BPOWer:BURSt:WIDTh?
[:SENSe]:BPOWer:BURSt:AUTO OFF|ON|0|1
[:SENSe]:BPOWer:BURSt:AUTO?
Remote Command Notes: Burst width cannot be set remotely as a percent of the current
burst width. That functionality is available from the front panel, but the
command must be sent with a time value.
Example:
SENS:BPOW:BURST:AUTO ON
5.8.7
Optimize Ref Level
Sets the input attenuator to optimize the robustness of the measurement, which is its
freedom from errors due to input compression.
NOTE
There will always be a minimum of 6 dB of attenuation set to protect the
analyzer input.
Key Path:
Meas Setup
State Saved:
Saved in instrument state.
Factory Preset: n/a
History:
Added with firmware revision A.02.00.
Remote Command:
[:SENSe]:POWer[:RF]:RANGe:AUTO ONCE
Example:
POW:RANG:AUTO ONCE