Chapter 5
One-Button Measurement Functions
Marker (Complementary Cumulative Distribution Function—CCDF)
O
n
e-
But
ton
M
easur
e
m
e
nt
Funct
ions
5.13.1
Select Marker
Selects one of the four possible markers. Once a marker is selected, it can be set to any of
the control modes
Normal
,
Delta
, or
Off
.
Key Path:
Marker
(after
MEASURE
,
Power Stat CCDF
is selected.)
State Saved:
The number of the selected marker is saved in instrument state.
Factory Preset: Marker 1
Remote Command:
:CALCulate:PSTatistic:MARKer[1]|2|3|4[:STATe] OFF|ON|0|1
:CALCulate:PSTatistic:MARKer[1]|2|3|4[:STATe]?
Sets or queries the state of a marker. Setting a marker to state ON or 1 selects that
marker. Setting a marker which is OFF to state ON or 1 puts it in
Normal
mode and places
it at the center of the display. Setting a marker to state OFF or 0 selects that marker and
turns it off. The response to the query will be 0 if OFF, 1 if ON.
Example:
CALC:PST:MARK2 ON
selects marker 2.
CALC:PST:MARK:STAT ON
will not modify a marker that is already on.
5.13.2
Normal
Sets the control mode for the selected marker to
Normal
. If the marker is off, a single
marker is activated at the 0 dB point of the selected trace (see
).
You can then adjust the selected marker using the numeric keypad, knob, or
Ý ß keys.
Key Path:
Marker
(after
MEASURE
,
Power Stat CCDF
is selected.)
Factory Preset: Off
Remote Command:
See
for the mode command.
Example:
CALC:PST:MARK:MODE NORM
selects marker 1 and sets it to
Normal
.
5.13.3
Delta
Sets the control mode for the selected marker to
Delta
. In
Delta
mode the display shows the
difference between the active (
Delta
) marker and a reference marker. When
Delta
mode is
selected the reference marker is placed at the current marker position, unless the marker
was
OFF
, in which case both the active marker and the reference marker are placed at 0 dB
point of the selected trace (see
). You can adjust the trace point of
the active delta marker. Annotation in the active function block and in the upper-right
corner of the display indicates the power ratio (dB) difference and probability (percentage)
difference of the two markers.