Agilent Technologies E4445A User Manual

Page of 664
394
Chapter 5
One-Button Measurement Functions
Marker (Complementary Cumulative Distribution Function—CCDF)
O
n
e-
But
ton
M
easur
e
m
e
nt
Funct
ions
Remote Command:
See
for the mode command.
Example:
CALC:PST:MARK3:MODE DELT
selects marker 3 as a delta marker and
places a reference marker at the 0 dB point on the specified trace (see
). If marker 3 is OFF it places both the active and
the reference markers at the 0 dB point of the specified trace.
5.13.4
Off
Turns off the selected marker. In addition,
Off
also turns off functions related to the
selected marker.
Key Path:
Marker
(after
MEASURE
,
Power Stat CCDF
is selected.)
Remote Command:
See
for the command to select the control mode.
Example:
CALC:PST:MARK3:STAT OFF
selects marker 3 and sets it to
Off
.
5.13.5
Marker Trace
Chooses which trace the selected marker will be placed on. You can pick
Measured
,
Gaussian
,
or
Reference
.
Key Path:
Marker
(after
MEASURE
,
Power Stat CCDF
is selected.)
State Saved:
The
Marker Trace
for each marker is saved in instrument state.
Factory Preset: Measured
Remote Command:
:CALCulate:PSTatistic:MARKer[1]|2|3|4:TRACe MEASured|GAUSsian|REFerence
Puts the marker on the specified trace.
:CALCulate:PSTatistic:MARKer[1]|2|3|4:TRACe?
The query returns the type of the trace on which the marker currently resides.
Example:
CALC:PST:MARK1:TRAC GAUSsian
places marker 1 on the Gaussian trace.
5.13.6
Marker All Off
Turns off all markers.