Agilent Technologies E4445A User Manual

Page of 664
Chapter 5
399
One-Button Measurement Functions
Meas Setup (Channel Power—CHP)
One-
But
ton
M
e
asur
e
m
e
nt
Funct
ions
5.14
Meas Setup (Channel Power—CHP)
When the channel power measurement has been selected in the
Measure
key menu of the
Spectrum Analysis Mode, this key displays the appropriate measurement setup menu.
The Channel Power measurement measures the power and power spectral density in the
channel bandwidth that you specify. One marker pair on the display indicates the edges of
the channel bandwidth. The center frequency, reference level, and channel bandwidth
must be set by the user.
Key Path:
Front-panel key
Dependencies/
Couplings:
Menu changes depending on the Mode and Measurement selected.
State Saved:
No save
History:
Added with firmware revision A.02.00
Remote Command:
There is no equivalent remote command.
5.14.1
Avg Number
To specify the number of measurement averages used when calculating the measurement
result set
Avg Number
to
On
. The average will be displayed at the end of each sweep. Setting
Avg Number
to
Off
disables measurement averaging.
Key Path:
Meas Setup
State Saved:
Saved in instrument state.
Factory Preset: 10 averages/ Off
Knob Increment: 1
Step Key
Increment:
1
Range:
1 to 1000
Remote Command:
[:SENSe]:CHPower:AVERage:COUNt <integer>
[:SENSe]:CHPower:AVERage:COUNt?
[:SENSe]:CHPower:AVERage[:STATe] OFF|ON|0|1
[:SENSe]:CHPower:AVERage[:STATe]?