Agilent Technologies E4445A User Manual

Page of 664
Chapter 5
401
One-Button Measurement Functions
Meas Setup (Channel Power—CHP)
One-
But
ton
M
e
asur
e
m
e
nt
Funct
ions
Factory Preset: 2 MHz, or as defined by the selected radio standard.
Terminators:
GHz, MHz, kHz, Hz
Default Terminator: Hz
Knob Increment: 1%
Step Key
Increment:
1,1.5, 2, 3, 5, 7.5, 10...
Range:
100 Hz to maximum span supported by your analyzer.
Remote Command:
[:SENSe]:CHPower:BANDwidth|BWIDth:INTegration <freq>
[:SENSe]:CHPower:BANDwidth|BWIDth:INTegration?
Example:
CHP:BAND:INT 1 MHz
CHP:BAND:INT?
5.14.4
Chan Pwr Span
Press
Chan Pwr Span
to set the analyzer span for the channel power measurement. When the
RRC Filter
is set to
On
, the lower limit for the span is
.
Key Path:
Meas Setup
State Saved:
Saved in instrument state.
Factory Preset: 3 MHz, or as defined by the selected radio standard.
Default Terminator: Hz
Knob Increment: 1, 1.5, 2, 3, 5, 7.5, 10...
Step Key
Increment:
1%
Range:
Current integration bandwidth to 10 times the integration bandwidth or
span of your analyzer.
Remote Command:
[:SENSe]:CHPower:FREQuency:SPAN <freq>
[:SENSe]:CHPower:FREQuency:SPAN?
Example:
CHP:FREQ:SPAN 2 MHz
CHP:FREQ:SPAN?
1
alpha
+
(
) Integration BW
×