Agilent Technologies E4445A User Manual

Page of 664
Chapter 2
89
Instrument Functions: A - L
BW/Avg
In
st
ru
m
e
nt
Funct
ions
:
A
-
L
Remote Commands:
[:SENSe]:AVERage[:STATe] OFF|ON|0|1
[:SENSe]:AVERage[:STATe]?
[:SENSe]:AVERage:COUNt <integer>
[:SENSe]:AVERage:COUNt?
[:SENSe]:AVERage:CLEar
Remote Command Notes: For valid average data, you must re-start the trace at the
beginning of a sweep. To do this remotely, first abort (:ABORT) the sweep
and then initiate a single sweep (:INIT:CONT OFF).
Example:
AVER ON
AVER:COUN 100
AVER:CLE
clears the current average and restarts the averaging process.
2.2.5
Avg/VBW Type
Displays the functions that enable you to automatically or manually choose one of the
following averaging scales: log-power (video), power (RMS), or voltage averaging.
NOTE
When you select log-power averaging, the measurement results are the
average of the signal level in logarithmic units (decibels). When you select
power average (RMS), all measured results are converted into power units
before averaging and filtering operations, and converted back to decibels for
displaying. Remember: there can be significant differences between the
average of the log of power and the log of the average power.
The following are the averaging processes within a spectrum analyzer, all of which are
affected by this setting:
Trace averaging (see
BW/Avg
) averages signal amplitudes on a trace-to-trace basis.
Average detector (see
Detector, Average
) averages signal amplitudes during the time or
frequency interval represented by a particular measurement point.
Noise Marker (see
Marker Noise
) averages signal amplitudes across measurement points
to reduce variations for noisy signals.
VBW filtering adds video filtering which is a form of averaging of the video signal.
When manual is selected, the type is shown on the left side of the display with a #. When
auto is selected, the analyzer chooses the type of averaging. When one of the average types
is selected manually, the analyzer uses that type regardless of other analyzer settings, and
sets
Avg/VBW Type
to
Man
.
Key Path:
BW/Avg
or
Auto Couple