Agilent Technologies E4445A User Manual

Page of 664
90
Chapter 2
Instrument Functions: A - L
BW/Avg
In
st
ru
m
e
nt
Funct
ions
:
A
-
L
Saved State:
Saved in Instrument State
Factory Preset: Auto (Log-power)
Remote Command:
[:SENSe]:AVERage:TYPE RMS|LOG|SCALar
[:SENSe]:AVERage:TYPE?
[:SENSe]:AVERage:TYPE:AUTO OFF|ON|0|1
[:SENSe]:AVERage:TYPE:AUTO?
Example:
AVER:TYPE:RMS
Sets Power (RMS) averaging
AVER:TYPE:SCAL
Sets Voltage averaging
AVER:TYPE:LOG
Sets Log-Power (video) averaging
2.2.5.1
Auto
Chooses the optimum type of averaging for the current instrument measurement settings.
Auto
selects Power (RMS) Averaging if
Marker Noise
is on,
Band/Intvl Power
is on, or
Detector
is
set to
Man
and
Average
. It selects Voltage Averaging if
Amplitude, Scale Type
is set to
Lin
. For
other conditions,
Auto
selects Log-Power Average.
Key Path:
BW/Avg
,
Avg/VBW Type
Remote Command:
See
Example:
AVER:TYPE:AUTO ON
2.2.5.2
Log-Pwr Avg (Video)
Selects the logarithmic (decibel) scale for all filtering and averaging processes. This scale is
sometimes call “Video” because it is the most common display and analysis scale for the
video signal within a spectrum analyzer. This scale is excellent for finding CW signals near
noise, but its response to noise-like signals is 2.506 dB lower than the average power of
those noise signals. This is compensated for in the Marker Noise function. When this type
of averaging is selected, LgAv appears on the left side of the display.
The equation for trace averaging on the log-pwr scale is shown below, where N is the
number of averages accumulated. (In continuous sweep mode, once N has reached the
Average Number, N stays at that value, providing a running average.)
Assumes all values in decibel scale.
New avg
N
1
(
)Oldavg Newdata
+
N
-------------------------------------------------------------------
=