Figure 13-20: Loops Display with Overlay Feature Active
Inspiratory Area
The Loops screen includes a window for the inspiratory area (Figure 13-21F).
This window displays a numeric value that represents the ventilator imposed
work of breathing, and is only calculated and displayed for a frozen loop. If the
ventilator is supporting the entire breath, the Insp Area window displays
“0.000”. If the Insp Area window shows dashes (“– – – –”), the ventilator is
not calculating inspiratory area because the loops are not frozen.
This window displays a numeric value that represents the ventilator imposed
work of breathing, and is only calculated and displayed for a frozen loop. If the
ventilator is supporting the entire breath, the Insp Area window displays
“0.000”. If the Insp Area window shows dashes (“– – – –”), the ventilator is
not calculating inspiratory area because the loops are not frozen.
Figure 13-21: Loops Screen with Inspiratory Area Window
SAVE button
OVERLAY icon
button
button
SINGLE VIEW
icon button
icon button
Dashes indicate
Inspiratory are
not calculated
(loop not frozen)
Inspiratory are
not calculated
(loop not frozen)