Figure 13-20: Loops Display with Overlay Feature Active
Inspiratory Area
The Loops screen includes a window for the inspiratory area (Figure 13-21F). 
This window displays a numeric value that represents the ventilator imposed 
work of breathing, and is only calculated and displayed for a frozen loop. If the 
ventilator is supporting the entire breath, the Insp Area window displays 
“0.000”. If the Insp Area window shows dashes (“– – – –”), the ventilator is 
not calculating inspiratory area because the loops are not frozen.