Philips V200 User Manual

Page of 328
Chapter 13
13-100
 
Respironics V200 Ventilator Operator’s Manual
REF 1057983 A
Options and Accessories
Initial Neonatal Settings
When the ventilator is first powered on and the neonatal patient type is 
selected, the initial startup settings found in Table 13-21: are applied. Each 
subsequent startup will display the settings from the last use of the ventilator 
of that patient type with the exception of O
2
. The last O
2
 setting used on the 
ventilator will remain regardless of patient type.
Initial Patient Settings, Ranges
Ventilation Settings
Value
Ranges
Ventilation Mode
PCV - SIMV
AC, SIMV, CPAP
Rate
20 BPM
1 - 150 BPM
Pressure
10 cmH
2
O
5 - 100 cmH
2
O
I-Time
0.3 Sec
0.1 - 2.0 Sec
PEEP
3 cmH
2
O
0  -  35  cmH
2
O
PSV
0 cmH
2
O
0 - 100 cmH
2
O
I-Trigger
2.0 L/Min
0.3 - 10.0 L/Min
E-Cycle
25%
10 - 80%
Rise-TIme
0.2 Sec
0.1 - 0.5 Sec
O
2
 %
same as last O
2
 setting
21 - 100%
Apnea Rate
20 BPM
1 - 80 BPM
Alarm Settings
Value
Ranges
High Press
30 cmH2O
10 - 105 cmH2O
Low Insp Press
5 cmH2O
3 - 105 cmH2O
Low Peep
0 cmH2O
0 - 35 cmH2O
Low Vt Mand
10 mL
0 - 300 mL
Low Vt Spont
3 mL
0 - 300 mL
High Rate
150 BPM
0 - 150 BPM
High VE
1.2 L
0.00 - 5.0 L
Low VE
0.3 L
0.00 - 5.0 L
Apnea
20 Sec
10 - 60 Sec
I-Time Too Long
sounds after 1.5 Sec of spont 
breath
Table 13-21: Initial Neonatal Patient Settings