Options and Accessories
REF 1057983 A
Respironics V200 Ventilator Operator’s Manual
13-101
Chapter 13
Percent Leak
When the neonatal software option is loaded on the ventilator, a % Leak 
window appears on the Patient Data screen. The window displays the 
estimated leak of the delivered volume as a percentage. The % Leak value is 
updated at the beginning of each inhalation. The % Leak is available for all 
patient types when the neonatal software option is loaded. While the estimate 
for percent leak is not a precise measurement it can be used to trend relative 
changes in leak delivered volume.
Figure 13-55: Patient Data Screen With % Leak
NOTE:
Patient data ranges and accuracies can be found in Chapter 12, 
“Technical Specifications” 
with the exception of the Exhaled Tidal 
Volume. The accuracy for exhaled Tidal Volume is ±(4ml ±10%)
NOTE:
Neither Respiratory Mechanics nor Expiratory Hold are active when the 
neonatal patient type is selected. Flow trigger is the only trigger type 
available. All other features apply.
Percent Leak