A&D GP-100K User Manual

Page of 98
 
 22
Using the accumulation function 
Use the keys below to operate the accumulation function. 
MODE key: 
Displays the weighing data and the total value alternately each time it is pressed. 
 
Will not change the unit while the accumulation function is in use. 
RE-ZERO key:  Sets the display to zero while the weighing data is displayed. 
 
Deletes the total value while the total value is displayed. 
PRINT key 
Outputs and adds the weighing data while the weighing data is displayed. 
 
Outputs the total value while the total value is displayed. 
 
 
1  Press the RE-ZERO key to set the display 
to zero. 
2  Place a sample on the pan. The weight 
value is displayed. 
3  Press the PRINT key. The weight value is 
added to the total and output.   
  The accumulation number at the upper left 
of the display increases by one. 
4  Repeat steps 1-3 to accumulate more data. 
5  Press the MODE key to display the total 
value. 
 
Outputting the value
 
6  Press the PRINT key to output the total 
value. 
7  Press the RE-ZERO key to delete the total 
value. 
 
 
 
Notes 
The output format depends on the function 
table setting. 
While the accumulation function is in use,   
the data memory function is not available. 
To disable the accumulation funciton, set the 
“Accumulation function (
add)” parameter to “0”. 
When the “Data number output (
d-no)” 
parameter is set to “
1”, the accumulation 
number will be output before the weighing 
data.