6.  RESPONSE ADJUSTMENT / SELF CHECK FUNCTION 
This function detects the influence on weighing that is caused by drafts and/or vibration at the place 
where the balance is installed and sets the response characteristic automatically. When this function 
is selected, the balance self-checks the performance at the same time.   
Two modes of response adjustment are available: automatic and manual. 
 
The function has three rates as follows:   
 
Changing the response rate changes the display refresh rate. 
Indicator  Parameter  Response characteristic 
Display refresh rate   
FAST 
Cond 0 
MID. 
Cond 1 
SLOW 
Cond 2 
Fast response, Sensitive value 
   
Slow response, Stable value 
If the response rate is changed as follows:
MID. or SLOW   FAST =10 times/second
FAST      MID. or SLOW = 5 times/second
 
Note 
To set the refresh rate of 5 times/second when the response rate is FAST or 10 times/second 
when the response rate is MID. or SLOW, change the “Display refresh rate (
5pd)” parameter 
of “Environment, Display (
ba5fnc)” in the function table. For details, refer to “9.    FUNCTION 
TABLE”. 
 
6-1  Automatic Response Adjustment / Self Check Function 
This function automatically updates the response adjustment by analyzing the influence of the 
environment on the weighing data and also self-checks the balance performance using the internal 
mass. 
Operation 
1  Press and hold the MODE key until RESPONSE is 
displayed, then release the key. 
2  The balance automatically starts to check the balance 
performance and sets the response characteristic.   
Caution  
Do not allow vibration or drafts to affect the balance 
during adjustment. 
3  After automatic adjustment, the balance displays the 
updated response indicator and returns to the 
weighing mode. The response indicator remains 
displayed for a while.
 
 e.g. 
“ 
MID. 
OK 
 
” 
  The example above indicates that the result of the self 
check is good and MID. is selected as the response 
rate.