7-4  Calibration Using an External Weight 
This function calibrates the balance using an external weight.   
 
Operation 
1  Plug in the AC adapter and warm up the balance 
for at least 30 minutes with nothing on the pan. 
2  Press and hold the CAL key until  Calout  is 
displayed, then release the key.   
3 The balance displays  Cal  0 . 
  If you want to change the calibration weight (a 
list of usable weights is shown on page 25), 
press the SAMPLE key and proceed to step 4. 
   If you use the calibration weight value stored in 
the balance, proceed to step 5. 
4  Specify the calibration weight value as follows: 
SAMPLE 
key  To switch the display 
condition to: “All of the 
segments blinking” 
(calibration weight selection 
mode) or “The last two digits 
blinking” (value adjustment 
mode). 
RE-ZERO 
key  To select the calibration 
weight or adjust the value. In 
the value adjustment mode, 
-15 digits appear after +15 
digits. 
PRINT 
key 
To store the new weight 
value. Even if the AC 
adapter is removed, the data 
is maintained in non-volatile 
memory. 
CAL key 
To cancel the operation and 
return to Cal  0 .
 
Note 
Digit, when used for the GP series balance, 
indicates the smallest displayable weighing value.