A&D GP-100K User Manual

Page of 98
 
 26
Display 
 
This indicator means “the balance is measuring calibration data”. 
Do not allow vibration or drafts to affect the balance while this 
indicator is displayed.   
 
7-2  Automatic Self Calibration 
(Calibration due to changes in temperature)
 
This function automatically calibrates the balance when the balance detects an ambient temperature 
change. If GLP output is selected in the function table, the balance outputs the calibration report or 
stores the data in memory. Automatic self calibration functions even if the display is turned off 
(standby state). 
Caution 
If something is on the weighing pan, the balance judges that it is in use and does not perform 
automatic self calibration. To maintain the calibrated state, keep the weighing pan clear while 
not in use. 
 
The displays shown below are related to the automatic self calibration. 
Indicates that the balance detects a change in ambient 
temperature and automatic self calibration will start. If the balance 
is not used for a few minutes with this indicator blinking, the 
balance performs automatic self calibration. The blinking duration 
depends on the environment. 
Indicates that the balance is measuring calibration data. Do not 
allow vibration or drafts to affect the balance while this indicator is 
displayed. After calibration, the balance returns to indicate the 
previous display. 
Note 
The balance can be used while the indicator blinks. But, it is recommended that to maintain 
the accuracy, stop using the balance and confirm that there is nothing on the pan and allow 
the balance to perform self calibration.