Fisher 13-986-127G/GR/GA 13-986-127GHK User Manual

Page of 16
 
  
Isotemp
â Laboratory and Chromatography Refrigerators
Page 7 of 12
Part Number: 104416  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
Hold-Off Time 
 
 
 
 
 
 
 
 
 
 
 
Error Codes 
 
 
 
 
 
 
 
 
 
 
For Example: 
The display indicates 4°C but a reference thermometer in the 
refrigerator chamber indicates 6°C.  The operator presses and holds 
the Menu keypad for 5 seconds, then changes the display value from 
0 to +2 by using the UP arrow key.  Press the Menu key again.  Now 
the display indicates a chamber temperature of °6 C and the 
controller begins to cool (as long as 3-min. delay period has expired) 
to the desired temperature of 4°C as shown on the reference 
thermometer.   
 
Allow an additional 30 to 40 minutes for the refrigerator to again 
stabilize.  If the display is still inaccurate, repeat the calibration offset 
procedure. 
 
    CAUTION:  INCREASING the offset value DECREASES  
  the temperature at which the system controls.  If the 
offset is inadvertently set too high, the chamber
’s 
control temperature can fall below zero and cause 
evaporator  freeze-up.  This is caused by the 
evaporator temperature not rising above 0°C and not 
defrosting during the compressor “off cycle”.   
 
NOTE:  While in any of the controller mode setups (temp set, 
calibration offset or units), the controller will wait 15 seconds for 
a new value to be entered.  If there is no keypad operation within 
the 15-second time window, control will automatically revert to 
the temperature display mode and the Mode
 LED will turn off. 
 
 
 
 
Following the offset menu is the hold-off (Ho) menu.  The display will 
momentarily flash “Ho”, followed by a number.  The number shown is 
the delay time in minutes between compressor activations.  Use the 
UP/DOWN keypads to change the value.   
 
Increasing the hold-off time will allow  additional time for the 
evaporator to defrost during the off-cycle.  This can reduce the 
chance of evaporator freeze-up during times of high humidity.  
Pressing the Menu keypad while in the Ho mode, enters the hold-off 
time displayed, and returns controller operation to the temperature 
display mode. 
 
 
 
Error codes indicate when the controller is sensing a problem.  A 
description of each is given below.  See the Troubleshooting Table 
for additional information on error codes. 
 
E1 Open 
sensor. 
E2 
Under temperature.  Temperature at sensor is less than  
–36°C. 
E3 
Over temperature.  Temperature at sensor is greater than  
37°C.