Part Number: 104416  
 
 
 
 
 
 
 
 
Temperature 
Controller 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
Setting the Temperature 
 
 
 
 
 
 
 
Units Select 
 
 
 
 
 
 
 
Calibration Offset 
 
 
 
 
 
 
 
 
Begin operation by inserting the key into the key switch located on the 
header panel.  The key switch is packed inside the envelope, which is 
shipped in the refrigerator chamber.  Turning the key switch to the ON  
|
 ) position will energize the evaporator fans and the digital 
controller.   
 
 
The digital temperature controller is located on the left side of the 
header panel (See Figure 4).  When the unit is initially turned on, the 
display will indicate current chamber temperature.  The temperature 
units will be indicated by the °C or °F LED located just to the right of 
the temperature display.  The refrigerator is factory set at 4°C.  When 
the unit is first energized the evaporator fans will operate however it 
will not immediately cool.  A 3-minute compressor delay is 
programmed into the controller to provide sufficient time for the 
evaporator to defrost and to allow the system pressure time to 
equalize.  When the delay time is over, the Cool LED will illuminate, 
the compressor will run and the chamber will cool. 
 
NOTE:  The compressor requires a 3-minute delay time 
between activations.  The unit will not cool during this 
delay.  The delay is also activated when the controller 
is first energized. 
 
 
 
To change the set temperature, press and release the Menu keypad 
once.  The display will flash SP and the Mode LED will be illuminated.  
The last set temperature will then be shown in the display. 
To change the temperature, press the UP or DOWN arrow key.  The 
adjustable temperature range is 1 to 12°C (34 to 54°F). When the 
desired set temperature is displayed, press the Menu keypad to enter 
the set temperature and activate the Units Select menu. 
 
 
The second menu selects the units.  The refrigerator control displays 
temperature in °C or °F.  The factory setting is °C.  Starting from the 
temperature display mode (Mode LED off), press the Menu key twice.  
The current temperature units are displayed.  To switch between 
units, press the UP or DOWN arrow key then press the Menu key to 
select.  The unit LED will be displayed on the right of the temperature 
display.   
 
 
 
In the event the refrigerator needs to be calibrated, a simple routine is 
available to adjust the display and control point to a referenced 
standard.  To set a temperature offset, press and hold the Menu 
keypad for 5 seconds.  The display will flash oS followed by the last 
temperature offset value.  The factory setting is 0.  To change the 
offset value, press the UP or DOWN arrow key, then press the Menu 
key.  The value shown in the display will be added to the previous 
temperature reading.  The display will then flash Ho