Electrical 
Connection 
 
Fig. 3   Place shelves at desired height on four shelf supports. 
 
 
 
 
NOTICE:  Insufficient line voltage is often the 
cause of compressor start-up failure.  It is 
strongly recommended that a dedicated 
circuit, conforming to the National Electrical Code, 
Article 440, be used for powering the refrigerator. 
 
 
CAUTION: Be sure that the power supply is the same voltage that is 
specified on the refrigerator’s data plate.  
 
The frequency and nominal voltage requirements for the unit are 
specified on the data plate, which is located on the interiors upper left 
side.  Only plug the unit into a power source that meets these 
requirements.  Low line voltage is often the cause of service 
complaints.  With the unit running, check that the line voltage is 
within ±10% of that specified on the data plate. 
 
 
WARNING: For personal safety this unit must be 
properly grounded. 
 
 
The power cord of this instrument is equipped with a three prong 
(grounding) plug (NEMA 5-15P).  This plug mates with a standard 
three prong (grounding) wall receptacle (NEMA  5-15R) to minimize 
the potential of an electrical shock hazard.  Chromatography models 
are equipped with a (NEMA 5-20P).  This plug must mate with a 
(NEMA  5-20R).