 27 − 
 
6 F 8 A 0 9 3 4  
 
 
CAUTION
 
(1)  Select the cable run location so they are away from electrical equipment (motors, transformers, 
or radio transmitters) which causes electromagnetic or electrostatic interference
(2)  Deterioration of flowmeter circuit insulation occurs if the converter interior or cable ends get wet 
or humidified. This in turn causes malfunction of flowmeter or noise problemsAvoid a rainy 
day if the flowmeter is to be installed outdoors
. Even indoors, prevent water from splashing over 
the flowmeter. Try to finish the wiring as quickly as possible 
(3)  The converter has a surge arrestor/protector installed inside. Therefore, do not conduct a 
withstand voltage test for the converter. To check the insulation of the converter, use a voltage 
of 250Vdc or less
(4)  After wiring, be sure to install the terminal block protection cover. 
 
5.1 Cables 
Use the kind of cables shown in Table 5.1 to wire the converter. 
Table 5.1 Installation Cables 
Name Cable 
name 
Nominal 
cross-sectional 
area 
Finished 
outer 
diameter 
Description 
Power cable 
3-core vinyl sheathed cable or 2-core vinyl 
sheathed cable 
2 mm² 
11~13mm 
CVV JIS C 
3401,IEC60695,IEC607
54,IEC60227,IEC60245 
or equivalent 
Output signal 
cable 
The number of conductors the cable contains differs depending on the 
specification of the output signal cable. 
Use a shielded cable of finished outer diameter 11 to 13mm and nominal 
cross-sectional area 1.25mm
2
CVV-S JIS -258-C or 
equivalent