Tektronix AFG3000 User Manual

Page of 230
Reference
2-12
AFG3000 Series Arbitrary/Function Generators Reference Manual
Output Menu
Table 2-9 shows the Output Menu. To access the Output menu, push the 
front-panel Top Menu button, and then push the bottom bezel button.
Table 2-9: Output Menu
First Level
Second Level
Description
Load Impedance
50 
Ω
Option buttons. Set the load impedance to 50 
Ω.
Load
1
1.   Range: 1 
Ω
  to 10 k
Ω.
When dBm is currently selected for output amplitude units, the amplitude units setting 
is automatically changed to V
p-p
 if you select high impedance.
Option buttons & Numeric input.
You can set the load impedance to any value 
from 1 
Ω to 10 kΩ.
High Z
Option buttons. Select high impedance.
Invert
Off  On
Selects Invert On to invert a waveform.
Limit
High Limit
Numeric input.
Specifies the higher and lower amplitude limit. 
See Level Meter on page 1-12.
Low Limit
Noise
Noise Add
Off  On
Selects Noise Add On to add the internal noise 
to a waveform.
Noise Level
2
2.   Range: 0% to 50% of amplitude setting (V
p-p
) of signal waveform, resolution 1%
The maximum value of noise range is limited by waveform and amplitude range. 
Sets the noise level.
External Add
3
Off  On
3.   AFG3100 and AFG3200 series only.
Selects External Add On to add an external 
signal to the CH 1 output signal.