Syntax and Commands
AFG3000 Series Arbitrary/Function Generators Reference Manual
3-17
CALibration[:ALL]
The CALibration[:ALL] command performs an internal calibration.
The CALibration[:ALL]? command performs an internal calibration and returns 
0 (Pass) or a calibration error code.
Group
Calibration and Diagnostic
Related Commands
*CAL?
Syntax
CALibration[:ALL]
CALibration[:ALL]?
Arguments
None
Returns
<NR1>
Examples
CALibration:ALL
performs an internal calibration.
CALibration:ALL?
performs an internal calibration and returns results. For example, it might return 0, 
which indicates that the calibration completed without any errors.
*CLS (No Query Form)
This command clears all the event registers and queues, which are used in the arbi-
trary/function generator status and event reporting system.
Group
Status
Syntax
*CLS
Arguments
None
Examples
*CLS
clears all the event registers and queues.
NOTE. 
The self-calibration can take several minutes to complete. During this time, 
the arbitrary/function generator does not execute any commands. Do not power off 
the instrument during the self-calibration.