Tektronix AFG3000 用户手册
Syntax and Commands
AFG3000 Series Arbitrary/Function Generators Reference Manual
3-17
CALibration[:ALL]
The CALibration[:ALL] command performs an internal calibration.
The CALibration[:ALL]? command performs an internal calibration and returns
0 (Pass) or a calibration error code.
0 (Pass) or a calibration error code.
Group
Calibration and Diagnostic
Related Commands
*CAL?
Syntax
CALibration[:ALL]
CALibration[:ALL]?
Arguments
None
Returns
<NR1>
Examples
CALibration:ALL
performs an internal calibration.
performs an internal calibration.
CALibration:ALL?
performs an internal calibration and returns results. For example, it might return 0,
which indicates that the calibration completed without any errors.
performs an internal calibration and returns results. For example, it might return 0,
which indicates that the calibration completed without any errors.
*CLS (No Query Form)
This command clears all the event registers and queues, which are used in the arbi-
trary/function generator status and event reporting system.
trary/function generator status and event reporting system.
Group
Status
Syntax
*CLS
Arguments
None
Examples
*CLS
clears all the event registers and queues.
clears all the event registers and queues.
NOTE.
The self-calibration can take several minutes to complete. During this time,
the arbitrary/function generator does not execute any commands. Do not power off
the instrument during the self-calibration.
the instrument during the self-calibration.