Adam Equipment 8141 User Manual

Page of 16
 
@Adam Equipment Company 2011                                                                            (P.N. 8141, C- 02/2011)
 
1.3 
WEIGHING  
To  weigh  using  the  TBB  balances  place  the  item  to  be  weighed  on  the  stainless  steel  pan.  The 
beam will move up. Adjust the weights on  the 3 beams until the beam pointer is aligned  with the 
zero mark again. Start with the largest weight and work towards the smallest.  Each weight should 
be  moved  until  the  beam  falls  below  zero  then  moved  back  one  notch  and  then  the  next  weight 
moved.   
 
The last weight moved will be the sliding weight on the front beam.  Slide this weight until the beam 
pointer  aligns  with  the  zero  mark.  Read  the  weight  by  adding  the  values  for  each  weight.  For 
example 400+20+5.7 = 425.7g 
 
If the sample is more than 610 grams it will be necessary to add a weight to the hanger on the right 
side of the beam.  Add the 500g counter weight first then increase to 1kg, 1kg plus 500g or 2 of the 
1kg masses.  Add these counter masses to  the  readings from the  sliding  beams to  determine  the 
mass.   Counter weights are supplied only with the TBB 2610S and TBB 2610T. 
 
 
USING A CONTAINER FOR THE SAMPLE,   WITH THE TBB 2610S BALANCE  
If an empty beaker is placed on the scale you must first determine the weight of the empty beaker.   
Then add the material to the beaker and weigh again.  The weight of the material within the beaker 
is the total weight minus the weight of the beaker when empty. 
 
USING A CONTAINER FOR THE SAMPLE,   WITH THE TBB 2610T BALANCE 
The  tare  mass  on  the  back  of  the  beam  can  be  used  to  counteract  the  weight  of  the  empty 
container,  up  to  225  grams.    Set  all  the  weights  to  show  zero  on  the  beams.      When  the  empty 
beaker  is  placed  on  the scale  slide  the  tare mass  along  the  rod  until  the  beam  pointer  is  aligned 
with the zero mark. Then add the material to the beaker and weigh as normal.  The weight of the 
material within the beaker is the weight being shown by the sliding weights. 
1.4 
FILLING TO A KNOWN WEIGHT 
Place the empty beaker on the balance and determine the weight or zero the scale (TBB 2610T).  
Add the desired weight to this value to determine the final weight.  Set the weights on the beam to 
represent  this  final  value.  Pour  the  material  into  the  beaker  slowly.    As  the  desired  weight  is 
approached, the beam will rise.  Stop filling when the beam pointer is aligned with the zero mark. 
1.5 
DENSITY DETERMINATION 
The  TBB 2610S and  TBB 2610T   balances are supplied as standard with three legs allowing the 
balance to be raised about 185mm (7”) above the surface. There is also a small hook attached to 
the bottom of the TBB’s weighing mechanism which allows products to be suspended via string (or 
wire) into a solution in a beaker below. This allows you to first weigh the product on the stainless 
steel top pan (in air) and then suspend the product in a solution using the below hook (in liquid) to 
allow you to calculate the density of the product. The TBB 610S does not include the legs.  
 
 
 