10.0 CALIBRATION 
 
ACTION DISPLAYS 
Press the [Tare/Zero] key during 
the self-test at power on. The scale 
will ask to enter the password. 
 
 
“ Pi n” 
Default Password is 0000. 
Enter “0” four times. 
Password can be changed in 
technical parameters. 
Press [Print]
 
 
“ Pi n”     “ ----” 
Select the scale to be set up by 
using the [Local/Remote] key. 
Press the [Print] key to enter the 
technical section.  
 
“tEch”  “   LocAL”   “      “   
“tEch”  “rEmo tE”   “      “   
 
Use the [U.Wt.] to select the 
weighing unit to be used for setting  
the calibration for the scale. The 
arrow in the “Weight” window will 
indicate the unit selected. 
Press the [Print] key to continue. 
 
  
“tEch”  “   Uni t”   “      “   
 
 
 
You will enter the first parameter -
Calibration. Display will ask to 
unload any weight on the platform. 
Press the [Print] key to continue. 
“Un L o Ad” 
 
 
 
   
Enter the calibration weight to be 
loaded and press [Print]
 
Load the calibration weight onto 
the scale and press [Print] again. 
 
The Calibration is complete, the 
scale will run the self-test during 
which the weight should be 
removed.  
 
“SEL”    “ 0010”      
 
 
“Lo Ad”