11.0 ERROR CODES 
 
During the initial power-on testing or during the operation, the scale may 
show an error message.  The meaning of the error messages are described 
below. 
 
If an error message is shown, repeat the procedure that caused the 
message, such as turning the balance on, calibration or any other functions.  
If the error message is still shown then contact your dealer for further support.   
 
ERROR CODE 
DESCRIPTION 
POSSIBLE CAUSES 
Err  4 
Initial Zero is greater than the 
permissible value (typically 4% 
of maximum capacity) when 
power is turned on or when the 
[Zero] key is pressed, 
Weight on the pan when turning 
the scale on. 
 
Excessive weight on the pan 
when zeroing the scale. 
 
Improper calibration of the scale. 
 
Damaged load cell. 
 
Damaged Electronics. 
 
Err 5 
Keyboard error.   
Improper operation of the scale. 
 
Err 6 
A/D count is not correct when 
turning the scale on. 
Platform is not installed. 
 
Load cell may be damaged. 
 
Electronics may be damaged. 
 
FAIL H 
or
 FAIL L 
Calibration error 
Improper calibration. 
 
If the problem persists contact 
your dealer or Adam Equipment 
for assistance.