11.0 ERROR CODES 
 
During the initial power-on testing or during operation, the scale may show an 
error message.  The meaning of the error messages is described below. 
 
If an error message is shown, repeat the step that caused the message. If the 
error message is still shown then contact your dealer for support. 
  
ERROR CODE 
DESCRIPTION 
POSSIBLE CAUSES 
Err  4 
Initial Zero is greater than 
allowed (4% of maximum 
capacity) when power is turned 
on or when the [Zero/Enter] key 
is pressed. 
Weight on the pan when turning 
the scale on. 
 
Excessive weight on the pan when 
zeroing the scale. 
 
Improper calibration of the scale. 
 
Damaged load cell. 
 
Damaged Electronics. 
 
Err 5 
Keyboard error.   
Improper operation of the scale. 
 
Err 6 
A/D count is not correct when 
turning the scale on. 
Platform is not installed. 
 
Load cell is damaged. 
 
Electronics is damaged. 
Err 7 
Scale is not stable when setting 
the percentage weighing. 
The scale is not stable. 
 
Improper operation of the scale. 
 
FAIL H 
or
 FAIL L 
Calibration 
error. 
Improper calibration (should be 
within + 10% of the factory 
calibration). The old calibration 
data will be retained until the 
calibration process is complete. 
 
Err 9 
Scale is unstable. 
There  is  vibration or draft making 
the scale unstable. 
 
Electronics may be damaged. 
 
To view the A/D count, press [Zero/Enter] and [%] at the same time while the 
scale is performing the initial check at power-on.  Press [Zero/Enter] to return 
to normal weighing. 
 