Adam Equipment azplus 6 User Manual

Page of 32
© Adam Equipment Company 2007 
16
  
 5.3
 
CALIBRATION PROCEDURE
 
 
The scales are sealed to prevent unauthorised calibration.   
 
To calibrate the scales, access to the PCB inside is needed to place a jumper in K6 
position. This is done by removing the tamper-proof label placed on a hole at the bottom of 
the scale. Through the hole, pick up the jumper (the jumper will be there placed on one of 
the pins on the PCB) and place it in the desired K6 position. If the hole is not provided in 
your scale, the security seal has to be broken to remove the top cover and gain access to 
the main PCB. Replace the cover and install the screws before starting the calibration. 
 
 
6.0 SERVICE 
PARAMETERS 
6.1  USING “0000” TO ENTER THE CALIBRATION PARAMETER 
 
After the jumper has been placed in position K6, apply power to the scale.   
 
During the self-test press the [Tare] key. 
 
The display will ask for a code number, “
Pn - - - -
“ on the Weight Display.  
 
To carry out the Calibration, enter the number “
0000
” then press [Tare]. The display will 
show “
F1
” will be enabled. 
 
The display will show “
UnloAd
” asking you to remove any weight from the scale,  
   
The display will tell you to add weight to the scale by showing “
LoAd
” “    
15
” “
kilo
” 
 
Add the weight shown, wait for stability then press the [Tare] key. 
 
The display will show “
SPAn
” “
PASS
” if the calibration is OK. Or it will show “
SPAN
” “
FAIL
” if 
calibration could not be completed correctly. 
 
The scale will return to the parameter menu.