GE LTC MAP User Manual

Page of 68
 
3-6
 
Section 3:  Installation 
February 3, 2000 
70055MP Revision D 
Alarm Configuration Jumpers 
Jumpers JP18 and JP19 set the annunciator relay outputs to normal or 
inverted.  Jumper JP13 configures the 
Acknowledge Alarm 
button. Refer to 
Table 3-3 for jumper positions.  Refer to Figure 3-3 for jumper locations. 
NOTE: JP13 must always be installed. 
Table 3-3: Alarm Configuration 
Jumper 
Position 
Description 
JP18 
INV 
NORM 
Annunciator 1 Output Inverted. 
Annunciator 1 Output Normal. 
JP19 
INV 
NORM 
Annunciator 2 Output Inverted. 
Annunciator 2 Output Normal. 
JP13 
Installed 
Removed 
Configures the Alarm Acknowledge button for use. 
Not applicable. 
 
Ground Jumpers 
Jumpers JP10 and JP14, when installed, are set to earth ground.  These 
jumpers must always be installed for accurate and safe operation. 
WARNING: Jumpers JP10 and JP14 must always be installed for accurate 
and safe monitor operation.