Special Tests
66
• Touch a drop of linearity solution to the 
center of the pink test square
If the white pad becomes completely 
saturated, you have applied too much 
linearity solution. Repeat the application 
with a new test strip.
• Check the confirmation dot on the back of 
the test strip to verify it has turned 
completely blue, indicating an adequate 
amount of solution was applied. 
If white patches or streaks are visible, you 
have not applied enough linearity solution. 
Repeat the application with a new test strip.
• Wipe the linearity solution vial tip with a 
lint-free cloth or tissue before replacing the 
cap.
Firmly insert the test strip all the way into the 
test strip holder until it stops (the 
confirmation dot should be facing down).