Lifescan surestep pro User Manual

Page of 117
79
If you cannot resolve a problem, call LifeScan Healthcare Professional Line: 
1 800 524-7226 (USA)
1 888 353-0800 (Canada)
Error Messages 
ERROR
POSSIBLE
CAUSE
SOLUTION
Error 1
Sample 
Application
The bedside unit 
detected a problem 
with the color of 
the test strip.
1.Repeat test with new test 
strip and:
• Apply blood to pink test 
square only.
• Make sure confirmation 
dot turns completely blue.
• Insert test strip within 
2 min of applying blood.
2. Check that sample hemat-
ocrit is within range: adult 
25–60%, neonatal 25–65%.
3.If you continue to get any 
error message, obtain a 
laboratory result, and 
• Compare color of test 
strip confirmation dot to 
Color Chart on test strip 
bottle to see if glucose 
level is very high or very 
low.
• Take appropriate steps 
to monitor patient and 
provide medical care while 
awaiting laboratory result. 
4.Clean bedside unit and 
repeat test with new test 
strip.
5.Contact LifeScan for help.
BU.book  Page 79  Monday, October 28, 2002  11:14 PM