National Instruments NI USB-6009 User Manual

Page of 32
© National Instruments Corporation
17
NI USB-6008/6009 User Guide and Specifications
Analog Input
You can connect analog input signals to the NI USB-6008/6009 through 
the I/O connector. Refer to Table 6 for more information about connecting 
analog input signals.
Analog Input Circuitry
Figure illustrates the analog input circuitry of the NI USB-6008/6009.
Figure 7.  Analog Input Circuitry
MUX
The NI USB-6008/6009 has one analog-to-digital converter (ADC). 
The multiplexer (MUX) routes one AI channel at a time to the PGA.
PGA
The progammable-gain amplifier provides input gains of 1, 2, 4, 5, 8,
10, 16, or 20 when configured for differential measurements and gain 
of 1 when configured for single-ended measurements. The PGA gain is 
automatically calculated based on the voltage range selected in the 
measurement application.
A/D Converter
The analog-to-digital converter (ADC) digitizes the AI signal by 
converting the analog voltage into a digital code.
AI
+2.5 V
REF
39.2 k
Ω
127 k
Ω
30.9 k
Ω
PGA
Input Range
Selection
AI FIFO
ADC
MUX